Differential characteristic |
Basic Delivery Set |
Technical Characteristics
UCD-50 - New universal ultrasonic flaw detector for expert and industrial application
|
|
New universal ultrasonic flaw detector with colour TFT a matrix 640õ480 and magnificent speed is optimal a choice for
the expert testing. Powerful, easy and portable (weight only 2 kg with the batteries), in the ergonomic
shock-resistant case from ABS a plastic, the device establishes the new standard for modern digital devices of the
manual testing. Flaw detector UCD-50 allows measuring a thickness of products with the big accuracy, to deduce a signal
in a form A - and B-scan and possesses all functions on full documenting of the testing, with software UdReport.
The mode bending around a signal maximum is realized on the basis of special algorithm of device UD2V-P46.
Certificated of Gosstandard of Russia
RU.C.27.003.A ¹27966,
is registred in the National register of means
of measurements under ¹34900-07
also admitted to application in the
Russian Federation.
|
At starting |
Basic Delivery Set |
Technical Characteristics
- Display: Colour TFT 135 x 100 mm (640 õ 480 pixels)
- Amplifier: broadband 0.4 - 15 ÌHz
- TGC/DAC functions
- Work with curve amplitude-interval: measurement in dB echelop signal relative curve and operation LED
- Pulse repetition frequency: to 800 Hz
|
- Monitor gate: 2 independent monitor gates with individual logic
- À-, Â- scan
- Interface RS232, USB
- Mains power: LiOn batterie or
via external plug-in power unit.
- Maximal 8 hours from batteries
|
At starting |
Differential characteristic |
Technical Characteristics
Basic Delivery Set
- UCD-50 main unit
- Batteries
- Power supply 220 V
- Cable SR50-Lemo 2 pieces.
- Four transducers
- Cable for connection PC
- Software support for connection PC
- Documentations
- Transport bag
Additional implements
- Protective bag
- Ultrasonic transducers for different application
- Standard samples
|
At starting |
Differential characteristic |
Basic Delivery Set
Technical Characteristics
- Calibration Range
- min.: 0 - 4 mks
max.: 0 - 1000 mks
adjustable with step 1 mks
- Pulse shift
- up to -0,5 mks to 996 mks
with step up to 0,02 mks
- The maximum length of a testinglable material
- to 3000 mm (echo-mode)
- Sound velocity range
- 1000 - 9999 m/s
- Proble delay
- 0 - 100 mks
with step 0,01 mks
- Damping
- 25 îhms / 50 îhms/ 1000 ohms
- Resiver
- 50 ohms / 600 ohms
- Probe pulse
- radio-frequency pulse, amplitude 50 or 200V
- Pulse repetition frequency
- to 800Hz (depending on parameters setting)/ 40 Hz
- Frequency range
- broadband 0.4-15 ÌHz (-6 dB)
- Controllable gain
- 110 dB, with step 0.5, 1, 2 or 6 dB
- Additional button +dB
- programming 0-40dB, with step 0,5dB
- TGC/DAC
- 90 dB, 12 dB/mks
curve recording with maximum of 10 reference
echoes
- Rectification
- RF(in all ranges),
full-wave, positive and negative half wave, B-scan
- Reject
- 0 - 90% screen height
- Monitor gate
- 2 independent monitor gates with individual logic,
sound indication
- Alarm
- luminous and sonorous
- Sound path measurement
- up to 0 to first signal in zone or between signals in zones
- Display of amplitude
- as a percentage of the screen height,
in dB to reference test echo,
in dB referens to DAC curve
- Display
- Colour, TFT 640 õ 480 pixels
130 õ 100 mm
- À-signal
- 480 x 300 pixels in a standard mode
640õ480 in full-screen mode
- Memory
- 100 datasets with À-signal
500 reports of the testing (signal, measured results, parameters transducers, date, time and name report)
- Interface
- RS232 / USB
- Dimensions transducers
- 2 Lemo0
- Batteries
- LiOn 8 À/hr
- Operating time
- 8 hours from batteries
- Mains power
- 220 V AC
- Operating temperature
- up to -20 C to +50 C
- Dimensions (H x W x D)
- 200 mm x 225 mm x 80 mm
- Weight
- 2 kg with batteries
|
At starting |
Differential characteristic |
Basic Delivery Set
|