Differential characteristic |
Basic Delivery Set |
Technical Characteristics
UD2V-P45 - the best choice for the expert and industrial testing
|
|
Universal ultrasonic flaw detector with possibilities of the analogue device
and advantages of digital processing of a signal.
It is intended for detection of defects (discontinuity and uniformity of materials)
in half-finished products, finished articles and welded joints, for measurement of depth
and their coordinates position, measurements of a thickness, measurement of velocity of propagation and attenuation of ultrasonic oscillations in a material.
Has the high-contrast liquid crystal display, it is simple and convenient in circulation.
Certificated of Gosstandard of
Russia RU.C.27.003.A №16022/1,
is registered in the National regiters
of means of measurements under №17498-03
also admitted to application in the
Russian Federation,
Kazakhstan,
Ukraine and
Republic of Belarus.
|
At starting |
Basic Delivery Set |
Technical Characteristics
- Range: up to 3 to 2975 mm (steel)
- Display: LCD 110 x 65 mm (240 х 128 pixels)
with adjustable backlight and contrast
- Amplifier: broadband 0.5 - 10 MHz,
the filtres 1.25, 2.5, 5 and 10 MHz
- Controllable gain: 110 dB
with step 0.5, 1, 2 and 6 dB
- TGC/DAC functions: 90 dB, 10 dB/mks
- Work with curve amplitude-interval: measurement
in dB echelop signal relative curve and operation LED
- Probing impulse: rectangular 200 V, with adjustable duration
- Pulse repetition frequency: to 800 Hz depending on parameters setting/40Hz
- Rectification: RF(in all ranges),
full-wave, positive and negative half wave
|
- Reject: 0 - 90% screen height
- Monitor gate: 2 independent monitor gate with individual logic,
sound indication
- "Freeze" А-signal
- Measurement of a thickness on one signal or between two signals
- Accuracy of indication of a thickness: 0.001 at thickness to 10 mm
- Memory datasets: 64 adjustments with the A-signal, with possibility
of input of the name of adjustment from keyboard
- Memory results: 10 files on 100 results of measurements
- Interface RS232: for transfer of A-range, parameters of work and results of measurements on PC
- Mains power: 4 NiMh batterie size "D"
or via external plug-in main unit
- 8 hours operating with backlight
- Weight: 2.5 kg with batteries
|
At starting |
Differential characteristic |
Technical Characteristics
Basic Delivery Set
- UD2V-P45 main unit
- AC-DC power supply 220V
- Probe cable CR 50-Lemo 2piece
- Two transducers
- Cable RS232 for connection PC
- Software support for connection PC
- Documentations
- Transport bag
Additional implements
- Protective bag
- Module for batteries size "D"
- Set batteries NiMh size "D" 9000 mA/hr
- Universal recharger
- Adapter circuit for connection to USB to port of the PC
- Ultrasonic transducers for diverse application
- Standard samples
|
At starting |
Differential characteristics |
Basic Delivery Set
Technical Characteristics the defectoscope
- Calibration Range
- min.: 0 - 2.9 mm (steel); 0 - 1 mks
max.: 0 - 2975 mm (steel); 0 - 1000 mks
- Sound velocity range
- 1000 - 9999 m/s
- Pulse shift
- up to -0,5 mks to 996 mks
- Pulse delay
- 0 - 100 mks
- Damping
- 50 ohms / 1000 ohms (to 25 ohms in simultaneos modes)
- Input impedance
- 50 ohms / 600 ohms
- Soud impulse
- quadrangular, amplitude 200 V,
with change interval up to 50 to 500 ns,
with stepm 25 ns
- Pulse repetition frequency
- to 800 Hz in a mode of the maximum frequency and 40 Hz
- Amplifier
- broadband 0.5-15 MHz (-3 dB)
the filtres 1.25, 2.5, 5 and 10 MHz
- Controllable gain
- 110 dB, with step 0.5, 1, 2 or 6 dB
- TGC/DAC
- 90 dB, 10 dB/mks
- Curve amplitude-distance (DAC)
- construction on 10 points, regulated on height
- Rectification
- positive or negative half wave,
complete, radiofrequency signal
- Reject
- linear, 0 - 80% screen height
- Monitoring gate
- two intependent monitor gates with individual logic,
sound indication
- Alarm
- luminous and sonorous
- Operating modes alarm
- defect in the first zone,
defect in the second zone,
defect in the first and in the second zones,
defect in one of zones,
on DAC - signal comparison
in the first zone from a curve amplitude-distance.
- Sound path measurement
- from 0 to the first signal in a zone or between
signals in zones, on front or on a signal maximum
- Measurement resolution
- 3.125, 6.25, 12.5 or 25 ns depending on PRF
- Display of amplitude
- as a percentage of the screen height,
in dB to reference test echo,
in dB reference to DAC curve.
- Display
- LCD with switchable backlight and contarst,
110 х 65 mm; 240 х 128 pixels
- А-signal
- 73 mm x 50 mm; 160 x 100 pixels
- Memory
- 64 datasets with А-signal
10 files on 100 results of measurements
- Interface
- RS232
- Dimensions transducers
- 2 SR50 or Lemo (in order)
- Batteries
- 4 NiMh size "D"
- Operating time
- 12-16 hours from batteries 9 А/hr
- Mains power
- external block 220 V AC
- Operating voltage
- 4 - 10 V DC
- Power consumption
- maximum 5 W, depending on
instrument setting
- Operating temperature
- up to -5 C to +55 C
- Dimensions (H x W x D)
- 160 mm x 225 mm x 45 mm
- Weight
- 2,5 kg with batteries
|
At starting |
Differential characteristic |
Basic Delivery Set
|