Differential characteristic | Basic Delivery Set | Technical Characteristics

UD2V-P45 - the best choice for the expert and industrial testing

flaw detector ud2v-p45  
Universal ultrasonic flaw detector with possibilities of the analogue device and advantages of digital processing of a signal.
It is intended for detection of defects (discontinuity and uniformity of materials) in half-finished products, finished articles and welded joints, for measurement of depth and their coordinates position, measurements of a thickness, measurement of velocity of propagation and attenuation of ultrasonic oscillations in a material.
Has the high-contrast liquid crystal display, it is simple and convenient in circulation.
Certificated of Gosstandard of Russia RU.C.27.003.A №16022/1, is registered in the National regiters of means of measurements under №17498-03 also admitted to application in the Russian Federation, Kazakhstan, Ukraine and Republic of Belarus.


Download operating manual
(file pdf, 500 Кb)


Price list for ultrasonic Flaw DetectorsPrice
(file xls, 274 Кb)

At starting | Basic Delivery Set | Technical Characteristics

  • Range: up to 3 to 2975 mm (steel)
  • Display: LCD 110 x 65 mm (240 х 128 pixels) with adjustable backlight and contrast
  • Amplifier: broadband 0.5 - 10 MHz, the filtres 1.25, 2.5, 5 and 10 MHz
  • Controllable gain: 110 dB
    with step 0.5, 1, 2 and 6 dB
  • TGC/DAC functions: 90 dB, 10 dB/mks
  • Work with curve amplitude-interval: measurement in dB echelop signal relative curve and operation LED
  • Probing impulse: rectangular 200 V, with adjustable duration
  • Pulse repetition frequency: to 800 Hz depending on parameters setting/40Hz
  • Rectification: RF(in all ranges), full-wave, positive and negative half wave
  • Reject: 0 - 90% screen height
  • Monitor gate: 2 independent monitor gate with individual logic, sound indication
  • "Freeze" А-signal
  • Measurement of a thickness on one signal or between two signals
  • Accuracy of indication of a thickness: 0.001 at thickness to 10 mm
  • Memory datasets: 64 adjustments with the A-signal, with possibility of input of the name of adjustment from keyboard
  • Memory results: 10 files on 100 results of measurements
  • Interface RS232: for transfer of A-range, parameters of work and results of measurements on PC
  • Mains power: 4 NiMh batterie size "D" or via external plug-in main unit
  • 8 hours operating with backlight
  • Weight: 2.5 kg with batteries

At starting | Differential characteristic | Technical Characteristics

Basic Delivery Set

  • UD2V-P45 main unit
  • AC-DC power supply 220V
  • Probe cable CR 50-Lemo 2piece
  • Two transducers
  • Cable RS232 for connection PC
  • Software support for connection PC
  • Documentations
  • Transport bag

Additional implements

  • Protective bag
  • Module for batteries size "D"
  • Set batteries NiMh size "D" 9000 mA/hr
  • Universal recharger
  • Adapter circuit for connection to USB to port of the PC
  • Ultrasonic transducers for diverse application
  • Standard samples

At starting | Differential characteristics | Basic Delivery Set

Technical Characteristics the defectoscope

Calibration Range
min.: 0 - 2.9 mm (steel); 0 - 1 mks
max.: 0 - 2975 mm (steel); 0 - 1000 mks
Sound velocity range
1000 - 9999 m/s
Pulse shift
up to -0,5 mks to 996 mks
Pulse delay
0 - 100 mks
Damping
50 ohms / 1000 ohms (to 25 ohms in simultaneos modes)
Input impedance
50 ohms / 600 ohms
Soud impulse
quadrangular, amplitude 200 V,
with change interval up to 50 to 500 ns,
with stepm 25 ns
Pulse repetition frequency
to 800 Hz in a mode of the maximum frequency and 40 Hz
Amplifier
broadband 0.5-15 MHz (-3 dB)
the filtres 1.25, 2.5, 5 and 10 MHz
Controllable gain
110 dB, with step 0.5, 1, 2 or 6 dB
TGC/DAC
90 dB, 10 dB/mks
Curve amplitude-distance (DAC)
construction on 10 points, regulated on height
Rectification
positive or negative half wave,
complete, radiofrequency signal
Reject
linear, 0 - 80% screen height
Monitoring gate
two intependent monitor gates with individual logic,
sound indication
Alarm
luminous and sonorous
Operating modes alarm
defect in the first zone,
defect in the second zone,
defect in the first and in the second zones,
defect in one of zones,
on DAC - signal comparison
in the first zone from a curve amplitude-distance.
Sound path measurement
from 0 to the first signal in a zone or between
signals in zones, on front or on a signal maximum
Measurement resolution
3.125, 6.25, 12.5 or 25 ns depending on PRF
Display of amplitude
as a percentage of the screen height,
in dB to reference test echo,
in dB reference to DAC curve.
Display
LCD with switchable backlight and contarst,
110 х 65 mm; 240 х 128 pixels
А-signal
73 mm x 50 mm; 160 x 100 pixels
Memory
64 datasets with А-signal
10 files on 100 results of measurements
Interface
RS232
Dimensions transducers
2 SR50 or Lemo (in order)
Batteries
4 NiMh size "D"
Operating time
12-16 hours from batteries 9 А/hr  
Mains power
external block 220 V AC
Operating voltage
4 - 10 V DC
Power consumption
maximum 5 W, depending on
instrument setting
Operating temperature
up to -5 C to +55 C
Dimensions (H x W x D)
160 mm x 225 mm x 45 mm
Weight
2,5 kg with batteries