Differential characteristic |
Basic Delivery Set |
Technical Characteristics
UD2V-P45.Lite - Modern the digital ultrasonic flaw detector with price like UD2-12/1
|
|
New digital ultrasonic flaw detector, created for replacement of out-of-date analog devices UD2-12.
Keeping opportunities of analog processing and having completely a digital path the device allows
carrying out all functions ultrasonic flaw detector UD2V-P45/46 except for preservation of results
of the testing and communication with the personal computer. Ultrasonic flaw detector can be modernized
in the further up to version UD2V-P46.
It is intended for detection of defects (discontinuity and uniformity of materials) in semi-finished items,
finished articles and welded joints, for measurement of depth and their coordinates, measurements of
thickness, measurement of velocity of propagation and attenuation of ultrasonic oscillations in a material.
Has the high-contrast liquid crystal display, it is simple and convenient in circulation.
Certificated of Gosstandart of Russia RU.C.27.003. A ¹16022/1, is registered in the National register
of means of measurements under ¹17498-03 also admitted to application in the Russian Federation,
Kazakhstan,
Ukraine and
Republic of Belarus.
|
At starting |
Basic Delivery Set |
Technical Characteristics
- Range: 3 to 2975 mm
- Display: LCD 110 mm x 65 mm (240 x 128 pixels) with switchable backlight
- Controllable gain: 110 dB
- TGC/DAC function
- Alarms for gates
- Pulse echo and dual operation
- PRF: up to 800 Hz to 40Hz
- Rectification: RF (in all ranges), full-wave, positive and negative half wave
- Reject: 0 to 80% screen height
|
- Two independent monitor gates with individual logic
- A-scan image freeze and zoom
- Trigonometric functions
- Digital measurement resolution: 0.001 mm up to 10 mm steel
- Memory: 12 datasets with A-scan and no measurements results
- Mains power: 4 NiMh batteries size "D" or via external plug in power unit
- 16 hours operating with backlight
- Weight: 2.5 kg with batteries
|
At starting |
Differntial characteristic |
Òechnical Characteristics
Basic Delivery Set
- UD2V-P45.Lite main unit
- AC-DC power supply 220V
- Probe cable CR 50-Lemo 2 pieces
- Two transducers
- Batteries set unit NiMh size "D" 9000 mA/hr
- Documentations
- Transport bag
Additional implements
- Ultrasonic transducers for differently application
- Standard samples
|
At starting |
Differntial characteristic |
Basic Delivery Set
Òechnical Characteristics
- Calibration range
- min.: 0 - 2.9 mm (steel); 0 - 1 mks
max.: 0 - 2975 mm (steel); 0 - 1000 mks
- Sound velocity range:
- 1000 - 9999 m/s
- Pulse shift
- up to -0,5 mks to 996 mks
- Probe delay:
- 0 - 100 mks
- Damping
- 50 ohms / 1000 ohms
- Resiver
- 50 ohms / 600 ohms
- Pulse intensity
- quadrangular, amplitude 200V,
with change interval up to 50 to 500 ns,
with step 25 ns
- Pulse repetition frequency
- up to 800 Hz or 40 Hz
- Amplifier
- broadband 1-10 MHz (-3 dB)
- Controllable gain
- 110 dB, with step 0.5, 1, 2 or 6 dB
- TGC/DAC
- curve recording with maximum of 10 reference echoes.
- Rectification
- RF (in all ranges), full-wave,
positive and negative half wave
- Reject
- linear, 0-80% screen height
- Monitor gate
- two independent monitor gates with individual logic, variable in steps of 1%
- Alarm
- sonorous and luminous
- Sound path measurement
- digital display of sound path between zero and the first echo signal
in the gate or between two echoes in the gates, measurement at echo flank
or peak with trigonometric flaw location.
- Measurement resolution
- 3.125, 6.25, 12.5 or 25 ns depending on PRF
- Display of amplitude
- as a percentage of the screen height
in dB to reference test echo
in dB reference to DAC curve.
- Display
- LCD with switchable backlight and
(110 mm x 65 mm; 240 x 128 pixels)
- À-signal
- 73 mm x 50 mm; 160 x 100 pixels
- Memory
- 12 datasets with À-signal
- Interface with PC
- no
- Dimensions transducers
- 2 SR50
- Batteries
- 4 NiMh size "C" or "D"
- Operating time
- 16 hours from batteries 9 À/hr "D"
with backlight
- Power supply
- external block 220 V AC
- Operating voltage
- 4 - 10 V DC
- Power consumption
- maximum 5 W, depending on instrument setting
- Operating temperature
- up to +5 C to +55 C
- Dimensions (H x W x D)
- 160 mm x 225 mm x 45 mm
- Weight
- 2,5 kg with batteries
|
At starting |
Differential characteristic |
Basic Delivery Set
|